General Specifications |
| Model: |
NPB-75 |
NPB-70 |
| Optional
Accessories |
Carrying
Case, Clamp, Communication Adapter Set, Calibration gas kit,
External Battery Charger, Extra Rechargeable battery pack,
battery pack carrying pouch |
| Display |
Graphic LCD
display (128 x 64) with LED backlight (contrast and
backlight control); two numeric fields, 3 digits each, using
7-segment LED; two-color LED alarm bar
|
| Displayed
Parameters |
|
CO2
and plethysmographic
waveforms, ETCO2,
SpO2,
pulse rate, respiration rate, trend data |
|
| CO2
waveforms and ETCO2,
respiration rate, trend data |
|
| Alarms |
| Adjustable
alarm limits for all parameters; audio level as defined
in ISO 9703-2 standard |
|
|
Power
Requirements |
|
AC input
of 100 to 230 VAC @ 60/50 Hz |
|
| Battery |
Rechargeable
NiMH; supplies power between 3 to 6 hours (depending on power
management). Fully recharges in approximately 4.5 hours |
| Dimensions |
8.11" H x 3.46" W x 2.06"
D
(206 mm x 88mm 53 mm) |
| Weight |
1.87
lb (850 g) |
1.66
lb (750 g) |
| Communication
Interface |
RS-232 interface @ 9600 baud; full
duplex serial interface via RJ-45 connector
|
| Environmental |
|
Storage |
-20°C to
55°C |
|
Operating
instrument/printer |
0°C to 55°C
0°C to 40°C |
|
| Operating
and Storage |
| Pressure |
430 mmHg to
795 mmHg |
| Altitude |
- 1,250 ft to
15,000 ft (-381m to 4,572m) |
| Humidity |
10% to 95%
noncondensing |
|
|
| Shock
test IEC 68-2-27 |
| Sinusoidal
vibration test IEC 68-2-6 |
| Random
vibration wide band test IEC 68-2-34 |
| Drop
test MIL-STD-810E, method 516.4-11 |
|
| Standards |
IEC
601-1/EN 60601-1, prEN 864, prEN 865, UL 544, UL 2601, EN
55011-Class B |
| |
|
Capnograph
Specifications |
| Display
Range |
0
to 99 mmHg |
| Sampling
Rate |
50
± 7.5 ml/min |
| Warm-Up
Time |
30
seconds typical, reaches ±5% steady-state accuracy within 3
minutes |
| Accuracy |
| 0 to 38
mmHg ± 2mmHg |
| 39 to 76
mmHg ± 5% of reading |
| 77 to 99
mmHg ± 8% of reading |
|
| Respiration
Rate |
0
to 150 breaths/minute |
|
|
|
Pulse
Oximeter Specifications |
| Display
Range |
| SpO2:
0% to 100% |
| Pulse Rate:
20 to 250 beats/minute |
|
| Accuracy
(%SpO2±1SD) |
| Adults: |
| 70% to
100% ± 2 digits |
| 0% to
69 unspecified |
|
| Neonates: |
70% to 100%
± 2 digits |
|
| |
|
CO2
Accessories |
| Filter
Line and FilterLine Set: |
24-hour
typical operation in an OR/EMS environment |
| Filter
Line H and FilterLine Set H: |
72-hour
typical operation in an ICU environment |
| Nasal
Filter Line: |
Nasal
Filter Line: Adult; pediatric and neonatal nonintubated
patients; 12-hour typical use |
| Microstream
Airway Adapters: |
Adult/pediatric
and infant/neonata |